Detection Efficiency of NaI(Tl) Detector in 511-1332 keV Energy Range


AKKURT İ. , Gunoglu K. , ARDA S. S.

SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS, 2014 (SCI İndekslerine Giren Dergi) identifier identifier

  • Cilt numarası:
  • Basım Tarihi: 2014
  • Doi Numarası: 10.1155/2014/186798
  • Dergi Adı: SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS

Özet

As it is important to obtain accurate analytical result in an experimental research, this required quality control of the experimental system. Gamma spectrometry system can be used in a variety of different fields such as radiation and medical physics. In this paper the absolute efficiency, peak to valley ratio, and energy resolution of a 3 '' x 3 '' NaI(Tl) detector were determined experimentally for 511, 662, 835, 1173, 1275, and 1332 keV photon energies obtained from Na-22, Mn-54, Co-60, and Cs-137 radioactive sources.